What does the "csv output error" message indicate on the PrimeTest 250+?

There is an issue with the csv download of the PT250+, this appears to be with the firmware version 2.70. To permanently fix this it will require a firmware update.

As a temporary measure you can manually correct the download data by:

Deleting “mA” in the “PE Leakage Unit” column

Deleting “Fail” in the “PE Leakage Result” column

The above should be done for both the IEC lead and rcd tests results.